Syringes
|
10, 20, 30, 50 mL of calibrated syringes
|
Flow rate
|
0.1-1500 mL/h (0.1 mL/h step)
|
50 mL syringe
|
0.1-900 mL/h (0.1 mL/h step)
|
30 mL syringe
|
0.1-600 mL/h (0.1 mL/h step)
|
20 mL syringe
|
0.1-400 mL/h (0.1 mL/h step)
|
10 mL syringe
|
Purge rate
|
1500 mL/h
|
50 mL syringe
|
900 mL/h
|
30 mL syringe
|
600 mL/h
|
20 mL syringe
|
400 mL/h
|
10 mL syringe
|
Accuracy of flow rate
(Nominal value)
|
Accuracy including syringe: within ±2%
Mechanical accuracy: within ±1%
|
Display range of volume delivered
|
0.1–9999 mL
(0.1–999.9 mL,0.1mL step)
(1000–9999 mL,1mL step)
|
Setting of delivery limit
|
0.1–9999 mL
(0.1–999.9 mL,0.1mL step)
(1000–9999 mL,1mL step)
|
Occlusion detection pressure
|
'H': 800±200 mmHg (106.7±26.7 kPa) or 1.09±0.27 kgf/cm2
'C': 500±100 mmHg (66.7±13.3 kPa) or 0.68±0.13 kgf/cm2
'L': 300±100 mmHg (40.7±13.3 kPa) or 0.41±0.13 kgf/cm2
|
Power
|
AC110~220V ±10%, 50~60 Hz ±1 Hz
Inner NiMH battery: DC12 V
|
Operation conditions
|
Ambient temperature: 5 to 40, Relative humidity: 20% to 80%
|
Storage conditions
|
Ambient temperature: -40 to 55, Relative humidity: â‰Â¤ 95%
|
Electric shock protection/class
|
Class I and internal power supply: Type CF, Continuous operation
|
Water proof level
|
IPX4
|
Weight
|
3.45kg (with clamp base)
|
Dimension
|
354mm (Width)×190mm (Height)×128mm (Depth)
|